CVE-2025-14096

Credential disclosure via insufficient operating-system protection

Published Dec 17, 2025 · Updated Dec 17, 2025

Credential disclosure in Radiometer ABL800, ABL90, and AQT90 FLEX analyzers allows physically proximate attackers to extract credentials. The customized Windows XP and Windows 7 operating systems insufficiently protect embedded credentials and execute with unnecessary privileges, but Radiometer has not disclosed the storage locus or extraction procedure. Physical access to an analyzer is required; extracted credentials expose authentication material, while Radiometer reports no public exploit code.

CVSS severity8.4
High
EPSS probability0.14%
Next 30 days · Sep 16, 2026
Known exploitationUnconfirmed
Based on sourced intelligence
Hinoki checkNot available
Coverage for this vulnerability

See if you're affected

Explore vulnerability checks for your environment with Hinoki.

Book a demo

Summary

What happened

Credential disclosure in Radiometer ABL800, ABL90, and AQT90 FLEX analyzers allows physically proximate attackers to extract credentials. The customized Windows XP and Windows 7 operating systems insufficiently protect embedded credentials and execute with unnecessary privileges, but Radiometer has not disclosed the storage locus or extraction procedure. Physical access to an analyzer is required; extracted credentials expose authentication material, while Radiometer reports no public exploit code.

The record

CVE
CVE-2025-14096
Published
Dec 17, 2025
Updated
Dec 17, 2025
Vendor
Radiometer Medical ApS
Product
ABL90 analyzer
Classifications
CWE-798, CWE-250, T1552
Attack vector
physical
Privileges
unauthenticated

Timeline

How it unfolded

  1. Dec 17, 2025CVE publishedPublication date reported by the CVE source.
  2. Dec 17, 2025Record updatedLatest update available in the CVE record.

Exploitability

Present is not the same as exploitable

Compare your product and version with the public record. A matching version still requires validation against your environment.

Is a vulnerable build present?

Compare these published version ranges with your installed build and any vendor patches.

  1. Affected versionversion=Windows 7 Operating system
  2. Affected versionversion=Windows XP Operating system

What conditions does exploitation require?

Attack vectorphysical
Required privilegesunauthenticated

What is affected?

Radiometer Medical ApS · ABL90 analyzerversion=Windows 7 Operating system; version=Windows XP Operating system
Radiometer Medical ApS · ABL800 analyzerversion=Windows 7 Operating system; version=Windows XP Operating system
Radiometer Medical ApS · AQT90 FLEX analyzerversion=Windows 7 Operating system; version=Windows XP Operating system

Published CVSS scores

8.4Radiometer Medical ApSCVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

CVSS describes severity. EPSS estimates exploitation probability.

Attacks

What attackers are doing with it

Daily unique IPs observed by Shadowserver honeypots for known exploited vulnerabilities (KEVs). Missing observations do not establish an absence of attacks.

Daily unique IPsNo honeypot observations are available for this CVE in the selected window.

No observations available

Sep 10, 2026Sep 16, 2026
Latest reporting daySep 16, 2026
Latest daily unique IPsUnavailable
Prior 30-day averageUnavailable
SourceShadowserver honeypots (KEV)
Vectorphysical
Privilegesunauthenticated
Known exploitationUnconfirmed
Public exploitUnconfirmed

Weakness, pattern, technique

CWE-798Use of Hard-coded Credentials
CWE-250Execution with Unnecessary Privileges
T1552Unsecured Credentials

Public exploit references

No public exploit references are available in this record.

Labels summarize the accepted research assessment. They do not indicate a test against your environment.

Technologies

Your stack

See the directory against your own environment.

Your stack

Check the software in your environment

Book a demo to see how Hinoki identifies affected software and validates exploitability in your environment.

Book a demo