CVE-2024-9832

Unauthorized device access via unlimited login attempts

Published Nov 14, 2024 · Updated Nov 18, 2024

Improper login throttling in Baxter Life2000 Ventilation System 06.08.00.00 and earlier allows physically proximate attackers to gain device access. The clinician-password handler accepts unlimited failed attempts for both the Clinician Password and Serial Number Clinician Password, enabling brute-force guessing. Physical access to an unattended device is required; successful login exposes device settings and data, and setting changes can interrupt ventilation.

CVSS severity9.3
Critical
EPSS probability0.18%
Next 30 days · Sep 16, 2026
Known exploitationUnconfirmed
Based on sourced intelligence
Hinoki checkNot available
Coverage for this vulnerability

See if you're affected

Explore vulnerability checks for your environment with Hinoki.

Book a demo

Summary

What happened

Improper login throttling in Baxter Life2000 Ventilation System 06.08.00.00 and earlier allows physically proximate attackers to gain device access. The clinician-password handler accepts unlimited failed attempts for both the Clinician Password and Serial Number Clinician Password, enabling brute-force guessing. Physical access to an unattended device is required; successful login exposes device settings and data, and setting changes can interrupt ventilation.

The record

CVE
CVE-2024-9832
Published
Nov 14, 2024
Updated
Nov 18, 2024
Vendor
Baxter International
Product
Life2000 Ventilation System
Classifications
CWE-307, T1110
Attack vector
local
Privileges
unauthenticated

Timeline

How it unfolded

  1. Nov 14, 2024CVE publishedPublication date reported by the CVE source.
  2. Nov 18, 2024Record updatedLatest update available in the CVE record.

Exploitability

Present is not the same as exploitable

Compare your product and version with the public record. A matching version still requires validation against your environment.

Is a vulnerable build present?

Compare these published version ranges with your installed build and any vendor patches.

  1. Affected versionversion=06.08.00.00 and prior

What conditions does exploitation require?

Attack vectorlocal
Required privilegesunauthenticated

What is affected?

Baxter International · Life2000 Ventilation Systemversion=06.08.00.00 and prior
Baxter International · Life2000 Ventilator Firmwareversion=0 <=06.08.00.00

Published CVSS scores

9.3BaxterCVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

CVSS describes severity. EPSS estimates exploitation probability.

Attacks

What attackers are doing with it

Daily unique IPs observed by Shadowserver honeypots for known exploited vulnerabilities (KEVs). Missing observations do not establish an absence of attacks.

Daily unique IPsNo honeypot observations are available for this CVE in the selected window.

No observations available

Sep 10, 2026Sep 16, 2026
Latest reporting daySep 16, 2026
Latest daily unique IPsUnavailable
Prior 30-day averageUnavailable
SourceShadowserver honeypots (KEV)
Vectorlocal
Privilegesunauthenticated
Known exploitationUnconfirmed
Public exploitUnconfirmed

Weakness, pattern, technique

CWE-307Improper Restriction of Excessive Authentication Attempts
T1110Brute Force

Public exploit references

No public exploit references are available in this record.

Labels summarize the accepted research assessment. They do not indicate a test against your environment.

Technologies

Your stack

See the directory against your own environment.

Your stack

Check the software in your environment

Book a demo to see how Hinoki identifies affected software and validates exploitability in your environment.