Network data interception via unauthenticated clear-text protocol
Published Jun 29, 2020 · Updated Aug 4, 2024
Clear-text communication in Baxter SIGMA SPECTRUM Infusion System 6.x and 8.x allows remote attackers to intercept operational data. The Wireless Battery Module exchanges infusion-delivery status and operational data with Baxter Gateway through an application-layer protocol that provides neither encryption nor authentication. Attackers must first gain access to the supporting network; exposed data excludes passwords and protected health information, and the reported effect is observation or man-in-the-middle interception.
Summary
What happened
Clear-text communication in Baxter SIGMA SPECTRUM Infusion System 6.x and 8.x allows remote attackers to intercept operational data. The Wireless Battery Module exchanges infusion-delivery status and operational data with Baxter Gateway through an application-layer protocol that provides neither encryption nor authentication. Attackers must first gain access to the supporting network; exposed data excludes passwords and protected health information, and the reported effect is observation or man-in-the-middle interception.
The record
- CVE
- CVE-2020-12040
- Published
- Jun 29, 2020
- Updated
- Aug 4, 2024
- Vendor
- Baxter International
- Product
- SIGMA SPECTRUM Infusion System
- Classifications
- CWE-319, T1557, T1040
- Attack vector
- network
- Privileges
- unauthenticated
Timeline
How it unfolded
- Jun 29, 2020CVE publishedPublication date reported by the CVE source.
- Aug 4, 2024Record updatedLatest update available in the CVE record.
Exploitability
Present is not the same as exploitable
Compare your product and version with the public record. A matching version still requires validation against your environment.
Is a vulnerable build present?
Compare these published version ranges with your installed build and any vendor patches.
- Affected versionversion=Sigma Spectrum v6.x model 35700BAX, Baxter Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module v9,11,13,14,15,16,v20D29,v20D30,v20D31,v22D24, Baxter Spectrum v8.x with Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum LVP v8.x w/Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24
What conditions does exploitation require?
What is affected?
Attacks
What attackers are doing with it
Daily unique IPs observed by Shadowserver honeypots for known exploited vulnerabilities (KEVs). Missing observations do not establish an absence of attacks.
Weakness, pattern, technique
Public exploit references
No public exploit references are available in this record.
Labels summarize the accepted research assessment. They do not indicate a test against your environment.
Technologies
Your stack
See the directory against your own environment.
Your stack
Check the software in your environment
Book a demo to see how Hinoki identifies affected software and validates exploitability in your environment.
Book a demo